Bandwidth : DC to 150 MHz
Attenuation : 10X
Input Resistance : 10 MΩ
Input Capacitance : approx. 9.5pF
Scope Input Resis : 1 MΩ
Compensation Range : 15pF to 24pF
Readout Pin : No
Ground Ref : Yes
ID Button : No
Max Rated Voltage : 500 V (DC + Peak AC)
|
 |
453, 453A, 454, 454A, 10A2A
The P6054A Probe is a miniature, passive, fast-rise, 10X attenuation probe designed primarily for use with
Tektronix oscilloscopes, such as 453, 453A, 454, 454A, and the 10A2A Plug-In Unit. The probe can be compensated
for use with other oscilloscopes or plug-in units that have an input capacitance of 15 to 24 pF (paralleled by
1 MΩ).
The probe consists of a small-diameter probe body assembly (especially useful in compact circuitry), a
3.5-foot, 6-foot, or 9-foot cable, and a compensating box with a BNC connector.
A ground reference pushbutton on the body of the probe permits the user to obtain a ground reference or
to determine which trace of a multitrace display includes the signal from the P6054A Probe.
The compensating box houses a compensation network that provides optimum transient response when the
probe is used with wideband oxcilloscopes. The probe can be low-frequency compensated to match the input
of the associated instrument by adjusting the variable capacitor through the hole in the compensating box
housing. Internal calibration adjustments provide high-frequency compensation. A snap-on cover gives access
to the internal adjustments.
(excerpts from P6054A Instruction Manual 010-6054-13)
Electrical
| Attenuation : | 10X within 3% (including 1 MΩ, ±2% amplifier input) |
| Input Resistance : | 10 MΩ within 2% (includes input R of amplifier). |
| Input Capacitance : |
3.5-foot cable : 9.5 pF, ±0.5 pF 6-foot cable : 12.5 pF, ±0.5 pF 9-foot cable : 13.5 pF, ±0.5 pF |
| Compensation Range : | 15 pF or less to at least 24 pF. |
| Bandwidth (-3 dB) : | 3.5-foot and 6-foot cables, @150 MHz 9-foot cable, @115 MHz. |
| Maximum Input Voltage : | 500 V (DC + peak AC), derated with frequency.
|
|
Physical
| Net Weight : | With 3.5-foot cable, @5 ounces (141.8 grams) With 6-foot cable, @5.25 ounces (148.8 grams) With 9-foot cable, @5.5 ounces (155.9 grams) |
| Dimension: |
PROBE BODY : @3.7 inches (9.4 cm), length;
@0.45 inch (1.1 cm), maximum outside diameter.
CABLE: @3.5 feet (1.1 meter), @6 feet (1.8 meters), or @9 feet (2.7 meters) between strain relief bases.
COMPENSATION BOX: Length, @1.7 inches (4.3 cm); Width, @0.6 inch (1.5 cm); Height, @1.2 inches (3.0 cm).
|
Environmental
Probe will operate within specifications over the following ranges:
Temperature : -15° to +75°C.
Altitude : To 15,000 feet.
|
(excerpts from P6054A Instruction Manual 010-6054-13)
| Tektronix Catalog Cost | Low Bid | High Bid | average |
|---|
| $ ? |
. | . | . |
|
$32.00 3/7/03 #2512519822 erlyman (new in bag - 2 bids $30 start)
|
|